Displayed values
Value
Pressure sensor at air outlet
Mask pressure
Flow derived values
Leak
Tidal volume
Respiratory rate
Minute ventilation
Value
Pressure measurement
Mask pressure
Flow derived measurements
Leak
2
Tidal volume
2,3
Respiratory rate
2,3
Minute ventilation
Results are expressed at ATPD (Ambient Temperature and Pressure, Dry)
1
Accuracy may be reduced by the presence of leaks, supplemental oxygen, tidal volumes <100 mL or minute
2
ventilation <3 L/min.
Measurement accuracy verified as per ISO 10651-6:2004 for Home Care Ventilatory Support Devices (Figure 101 and
3
Table 101) using nominal ResMed mask vent flows.
32
Range
4–20 cm H
4–25 cm H
0–200 L/min
0–4000 mL
0–50 BPM
0–30 L/min
Accuracy
1
±0.5 cm H
1
±12 L/min or 20% of reading, whichever is greater, at 0 to 60 L/min
±20%
±1.0 BPM
±20%
2,3
O (CPAP);
2
O (ASV)
2
1
O (+ 4% of measured value)
2
Display resolution
0.1 cm H
O
2
1 L/min
1 mL
1 BPM
0.1 L/min