Displayed Values - ResMed H5i Clinical Manual

S9 series vpap adapt positive airway pressure device heated humidifier
Hide thumbs Also See for H5i:
Table of Contents

Advertisement

Displayed values

Value
Pressure sensor at air outlet
Mask pressure
Flow derived values
Leak
Tidal volume
Respiratory rate
Minute ventilation
Value
Pressure measurement
Mask pressure
Flow derived measurements
Leak
2
Tidal volume
2,3
Respiratory rate
2,3
Minute ventilation
Results are expressed at ATPD (Ambient Temperature and Pressure, Dry)
1
Accuracy may be reduced by the presence of leaks, supplemental oxygen, tidal volumes <100 mL or minute
2
ventilation <3 L/min.
Measurement accuracy verified as per ISO 10651-6:2004 for Home Care Ventilatory Support Devices (Figure 101 and
3
Table 101) using nominal ResMed mask vent flows.
32
Range
4–20 cm H
4–25 cm H
0–200 L/min
0–4000 mL
0–50 BPM
0–30 L/min
Accuracy
1
±0.5 cm H
1
±12 L/min or 20% of reading, whichever is greater, at 0 to 60 L/min
±20%
±1.0 BPM
±20%
2,3
O (CPAP);
2
O (ASV)
2
1
O (+ 4% of measured value)
2
Display resolution
0.1 cm H
O
2
1 L/min
1 mL
1 BPM
0.1 L/min