ResMed H5i Clinical Manual page 21

S9 series vpap adapt positive airway pressure device heated humidifier
Hide thumbs Also See for H5i:
Table of Contents

Advertisement

Parameter
Modes
CPAP
Access
ü
Date
ü
Time
ü
Reminders
Mask
ü
Water tub
ü
Tube
ü
Filter
ü
SD card
ü
Service
ü
Customised
ü
messages
(Custom 1,
Custom 2)
*[
Asia Pacific]
Default
Range
ASV
Full
Full / Limited
ü
DD Mmm YYYY
ü
00:00 (24 hr)
ü
6
Seven-day
ü
[12]*
increments
(starting from the
current set date)
6
ü
with a recurrence
[Off]*
period of one to 24
months.
6
ü
[Off]*
6
ü
Off
ü
Off
ü
[24]*
Off
ü
Description
Enables or disables S9
Essentials—If set to Limited, the
Info and Setup menu buttons
are disabled. This means that
the patient can simply start or
stop therapy and adjust ramp,
humidification or Climate Control.
Combined button presses remain
enabled.
Sets the current date or time. If
you set a new date or time that
occurs in the past then an 'Invalid
date/time, data exists for this
period' message is displayed.
Before this change can be made,
erase the compliance data –
available under the Configuration
menu.
A timed reminder to remind
a patient when they need to
replace their mask.
A timed reminder to remind
a patient when they need to
replace their water tub.
A timed reminder to remind
a patient when they need to
replace their tubing.
A timed reminder to remind a
patient when to replace the air
filter.
A timed reminder to remind a
patient that they need to remove
their SD card and return it to
you, enabling you to establish
compliance.
A timed reminder to remind a
patient when to return the device
for service.
Customised reminders, eg, to
return equipment or to phone a
particular person or number.
Custom reminder text can be
up to 16 characters long, via
a PC application. See your PC
application manual for more
information.
17