Indications for use
The H5i is indicated for the humidification of the air delivered from a CPAP device. The H5i is for use only as
recommended by a physician. The H5i is intended for single patient re-use in the home environment and re-use in a
hospital/institutional environment.
Contraindications
The H5i is contraindicated for use with patients whose upper (supraglottic) airway has been bypassed. Please refer to
the S9 Information Guide for contraindications associated with CPAP therapy.
Troubleshooting
If there is a problem, try the following suggestions. If the problem cannot be solved, contact your equipment supplier
or ResMed. Do not attempt to open these devices.
Problem/possible cause
No display
Power is not connected.
The S9 and H5i are not connected correctly.
Insufficient air delivered from the device
The S9 is not working correctly.
Air tubing is not connected properly.
Air tubing is blocked, pinched or punctured.
The H5i cover is not latched correctly.
The H5i cover seal is not fitted correctly.
Incorrect air tubing selected.
Solution
Ensure the power cable is connected and the power outlet (if available) is
on.
Ensure that the S9 and H5i are securely attached.
See the Troubleshooting section of the S9 Information Guide.
Check air tubing.
Unblock or free the air tubing. Check the air tubing for punctures.
Make sure the cover is latched onto the docking station correctly.
Make sure the cover seal is facing the right way up and fitted securely.
If you are using the SlimLine or Standard air tubing ensure that you have
the correct air tubing selected via the menu.