Page 1
This disinfection guide is intended for multipatient use of the H5i in a sleep lab, clinic, hospital or at a health care provider. If you are using the H5i as a single user in the home, refer to the Welcome Guide for cleaning instructions.
8. Locate the air outlet on the back of your H5i. 9. Remove the outlet clip by pulling apart the clip ends and sliding it off. 10. Remove the air outlet by pushing it through to the inside of the H5i. English...
Process Temperature Process Time 70 °C 100 minutes 75 °C 30 minutes 80 °C 10 minutes 90 °C 1 minute ResMed water tubs have been tested for disinfection for 100 cycles using hot water approximately 93°C for 10 minutes. English...
Sekusept aktiv™ 2.0% 30 minutes Sterilization The H5i cleanable water tub and air outlet have been validated for 100 cycles for sterilization by STERRAD™ 100 S, STERRAD 100 NX and STERRAD NX. Inspecting Perform a visual inspection of all components. If any visible deterioration is apparent (cracking, crazing, tears etc), the water tub should be discarded and replaced.
See www.resmed.com for other ResMed locations worldwide. For patent and other intellectual property information, see www.resmed.com/ip. H5i is a trademark of ResMed Ltd. CIDEX and Sterrad are trademarks of Johnson & Johnson. Alconox is a trademark of Alconox Inc. Neodisher MediZym is a trademark of Chemische Fabrik Dr Weigert.