Troubleshooting
1. Troubleshooting
If the instrument does not seem to be working properly, the user should first try to inspect and adjust it himself.
* If the user is unable to inspect the instrument himself or if the instrument fails to work properly after being adjusted or after
parts are replaced, contact your local dealer or J. MORITA OFFICE.
Problem
Check the battery power.
No power.
Check the battery installation.
Degraded battery.
Is there a sound when the instrument is
Display does not appear.
turned on and off?
Motor handpiece does not
Is it set for EMR mode?
run.
No sound.
Beep volume set to 0?
Beep sound an alarm even
Is the instrument set to CCW (reverse rota-
though the instrument is
tion) mode?
not being used.
Is contrary electrode properly hooked in the
corner of the patient's mouth?
Is the instrument set to EMR mode?
Is auto start turned off?
Motor does not run when
the file is inserted in the
canal.
Does the canal length indicator bar light up
only 1 bar or it does not light up?
Does the fixing screw for the built-in elec-
trode or external file electrode loose?
Does the canal length indicator bar light up?
Motor stops too easily.
Does the fixing screw for the built-in elec-
trode or external file electrode loose?
Torque limit may be set.
Motor spontaneously
Is the apical action setting on reverse?
starts running in reverse.
Is the instrument set to CCW (reverse rota-
tion) mode?
Torque limit value might be set too low.
The Apical Torque Down function might be
turned on.
Motor reverse its rotation
too easily.
Does the canal retain some blood or chemi-
cal solution?
Operation 2018-04-06
Check Points
Remedies
Charge the battery.
Install the battery properly.
Replace the battery.
Charge battery if there is no sound.
Broken display if there is a sound.
Select a mode other than EMR mode.
Set beep volume to 1, 2, or 3.
When set to CCW mode, the beeper sounds an alarm
after a set time period elapses. If this is annoying, set
the beeper sound for 0.
Hook the contrary electrode in the corner of the pa-
tient's mouth.
Select a mode other than EMR mode.
Turn the auto start function on.
Advance the file down the root canal, or add some
moisture such as saline solution to the canal to light up
2 bars or more.
Tighten the screw securely.
Advance the file down the root canal, or add some
moisture such as saline solution to the canal to light up
1 bar or more.
Tighten the screw securely.
Set the Torque Reverse function for R.L (torque reverse
less) if this is not desired.
Change the Apical Action setting to Off or Stop.
Change rotation mode to something other than CCW
(reverse rotation) mode.
Increase the torque limit value.
The torque limit automatically decreases as the file ap-
proaches the apex. To use a fixed reverse torque value,
turn the Apical Torque Down function off.
In that case, the canal measurement meter could indi-
cate a large movement and reach the flash bar.
Advance the file down the root canal so that the meter
display will resume in the appropriate position and the
file rotation will return to the forward direction.
Ref.
p.26
p.42
p.26
p.34
p.39
p.18
p.34
p.37
p.37
p.13
p.37
p.13
p.35
p.36
p.34
p.35
p.39
p.19
47