Tri Auto ZX2 Quick Guide

Before Use and Maintenance

Before Use

Component
Step 1
Assembly
Connect Contra Angle

Function Check

Step 1
Canal Measurement Function
Turn the power on. The LCD display shows the "m1" screen.
Touch the contrary electrode (hook) with the file and the clip
on the end of the file holder; make sure that all the indicator
bars on the meter in the LCD display light up.
If all the indicator bars do not light up, stop using the instru-
ment immediately and have it professionally repaired.
Turning the Power On  /   O ff
On
Off
2
1
Hold Down
Trademarks (™) and Registered Trademarks (®): The names of companies, services, etc. used in this manual are either trademarks or registered trademarks.
* Be sure to read the accompanying Instructions for Use before using the Tri Auto ZX2.
* Always wear appropriate gloves when you perform maintenance.
Step 2
Put on HP Protective Sleeve

Calibration

(Before first use after purchase, after replacing the contra angle, if
OTR action is triggered erratically or at random, etc.)
Connect the contra angle. With the instru-
ment turned off, hold down the
Left-Set switch (<) and then press
the Main switch.
Press the Right-Set switch (>) to start cali-
bration. After calibration, the instrument will
automatically return to the Standby display.
Step 3
Connect Probe
Cord
Be sure to connect the con-
trary electrode (hook) to the
white probe connector.
Step 2
Motor Rotation
Select "m4" (OTR mode).
Run the motor to see if it rotates smoothly.
• If the motor alternates between forward and
reverse rotation: Calibrate the instrument.
• If the instrument makes an abnormal vibration
or noise: Stop using the instrument immediately
and have it professionally repaired.
2
1
Hold Down
Step 4
Install File
2
1
Checking with Tester (once a week)
Turn the power on. Connect the tester.
Check the meter reading.
• If the meter lights up to within
two bars of bar number 1: Normal
• If the meter lights up to three bars
more or less than bar number 1:
Stop using the instrument immediately and have
it professionally repaired.
3

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the Tri Auto ZX2 and is the answer not in the manual?

Questions and answers

Subscribe to Our Youtube Channel

Summary of Contents for Morita Tri Auto ZX2

  • Page 1 Tri Auto ZX2 Quick Guide * Be sure to read the accompanying Instructions for Use before using the Tri Auto ZX2. * Always wear appropriate gloves when you perform maintenance. Before Use and Maintenance Before Use Component Step 1 Step 2...
  • Page 2 Cleaning Disinfection Lubrication Packing Sterilization (only for the contra angle) Sterilize the following: Recommendation: Spray the MORITA MULTI SPRAY a. File Holder into the connection end of the contra Drying Sterilizer Time time after b. Contrary Electrode angle for about 2 seconds.

Table of Contents