Manufacturer's Declaration; Electromagnetic Emissions; Electromagnetic Immunity - Somnetics Transcend Auto User Manual

Sleep apnea therapy system
Hide thumbs Also See for Transcend Auto:
Table of Contents

Advertisement

Transcend Auto User Manual

Manufacturer's declaration

This section presents the following topics:

Electromagnetic emissions

Electromagnetic immunity

IEC 60601-1 Compliance
Electromagnetic emissions
The Transcend Auto is intended for use in the electromagnetic environment specified
below. The customer or the user of the system should ensure that it is used in such an
environment.
Emissions test
RF radiated emissions
CISPR 11
RF conducted
emissions
CISPR 11
Harmonic emissions
IEC 61000-3-2
Voltage fluctuations/
flicker emissions
IEC 61000-3-3
Electromagnetic immunity
The Transcend Auto is intended for use in the electromagnetic environment specified
below. The customer or the user of the system should ensure that it is used in such an
environment.
Immunity test
Electrostatic
discharge (ESD)
IEC 61000-4-2
Electrical fast
transient/burst
IEC 61000-4-4
Surge
IEC 61000-4-5
Compliance
Electromagnetic environment—guidance
Group 1
The Transcend Auto uses RF energy only for its internal
function. Therefore, its RF emissions are very low and are
not likely to cause any interference in nearby electronic
equipment.
Class B
The Transcend Auto is suitable for use in all establishments,
including domestic establishments and those directly
connected to the public low-voltage power supply network
that supplies buildings used for domestic purposes.
Class A
Complies
IEC 60601 test
Compliance level
level
±2, 4, 6 kV contact
N/A. The Transcend Auto
does not have
conductive surfaces.
±8 kV air
±2, 4, 6, 8 kV air
±2 kV for power
±2 kV for power supply
supply lines
lines
±1 kV for
input/output lines
±1 kV for input/output
lines
±1 kV differential
±0.5*, 1 kV differential
mode
mode
±2 kV common
±2 kV common mode
mode
Electromagnetic
environment—guidance
Floors should be wood, concrete,
or ceramic tile. If floors are
covered with synthetic material,
the relative humidity should be
at least 30%.
Line power quality should be
that of a typical commercial or
hospital environment.
Line power quality should be
that of a typical commercial or
hospital environment.
Page 36