Transcend Auto performance

Working pressure range:
Accuracy of pressure setting:
Maximum system shutdown pressure:
Ramp time duration:
Operating temperature range:
Storage/transport temperature range:
Operating humidity range:
Storage/transport humidity range:
Altitude range:
4 to 20 cm H
O
2
±1 cm H
O or ±10%, whichever is greater
2
30 cm H
O
2
0-45 min + 25% time variance
41 to 95°F (5 to 35°C)
-4 to 140ºF (-20 to 60ºC)
10% to 80% relative humidity, non-condensing
10% to 90% relative humidity, non-condensing
0-8000 feet (Automatically adjusted)