Transcend 3 miniCPAP™
User Manual

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the Transcend 3 miniCPAP and is the answer not in the manual?

Questions and answers

Summary of Contents for Somnetics Transcend 3 miniCPAP

  • Page 1 Transcend 3 miniCPAP™ User Manual...
  • Page 2 Trademark Somnetics ® and Transcend ® are registered trademarks of Somnetics International, Inc.; all rights reserved. Copyright © Copyright 2019 Somnetics International, Inc.; all rights reserved. Contact Somnetics International, Inc. Corporate headquarters Somnetics International, Inc. Email info@somnetics.com http://www.mytranscend.com/ Telephone 651.621.1800 Toll-free telephone 877.621.9626...
  • Page 3 UL 60950-1 or IEC 60950-1 safety standards. • Do not plug any devices into the Transcend 3 miniCPAP USB port other than a computer that is compliant with 60950-1 safety standards. Attaching any other device to the Transcend 3 miniCPAP USB port may damage Transcend and may not be safe for the user.
  • Page 4: Table Of Contents

    Using the EZEX function .................. 26 Ending therapy ....................26 Drying mode ....................26 Replacing the Filter Media ................27 Caring for your Transcend 3 miniCPAP and components ........28 Cleaning the exterior ..................28 Cleaning of accessories………………………………………………………………………………………... 29 Cleaning for Multiple Users................29 Software User Guide Downloading device software…………………………………………………………………………………………31...
  • Page 5 AC power supply PSA2 ................... 44 Mobile power adapter (optional) — MPA1 ............44 Batteries (optional) — P and P Battery ............. 44 Transcend 3 miniCPAP performance ..............45 Manufacturer’s declaration ................46 Electromagnetic emissions................46 Electromagnetic immunity ................47 IEC 60601-1 compliance ................48...
  • Page 6: Introduction

    Transcend Auto User Manual Page 6 Introduction The Transcend 3 miniCPAP provides positive airway pressure to users in the range of 4 to 20 cm as prescribed by the clinician. The Transcend 3 miniCPAP product family includes the Transcend 3 miniCPAP and the Transcend 3 miniCPAP Auto. Buttons and LED lights facilitate control and provide operational feedback.
  • Page 7: Precautions For Use

    Indicates a tip, explanation or feature to aid in understanding, or efficient operation of the device. Warnings • Do not allow water to enter this device. Transcend 3 miniCPAP should not be exposed to environmental conditions where the system may get wet. • This device is not intended for life support.
  • Page 8: Cautions

    Discontinue use of the Transcend 3 miniCPAP and contact your physician if respiratory or skin irritations occur. • Do not introduce objects into the Transcend 3 miniCPAP air inlet or air outlet. • Inspect the power supply for signs of wear or damage before each use. Replace the...
  • Page 9 Transcend 3 miniCPAP User Manual Page 9 • Somnetics recommends replacing the air delivery tubing (hose) after every three months of use. • To protect the environment, some parts and accessories of the Transcend 3 miniCPAP, including optional batteries, must be disposed of in accordance with local regulations.
  • Page 10: Symbols

    Transcend 3 miniCPAP User Manual Page 10 Symbols Attention: Consult accompanying documents Protected against finger-sized objects and against dripping water when tilted 15 degrees from specified orientation IP22 Type BF Applied Part UL Seal of Approval demonstrating quality, safety and professional...
  • Page 11 Transcend 3 miniCPAP User Manual Page 11 Batch code Date of Manufacture Authorized Representative in the European Community Manufacturer Fragile, handle with care Keep dry Serial Number...
  • Page 12: Components Of The Transcend 3 Minicpap

    Transcend 3 miniCPAP User Manual Page 12 Components of the Transcend 3 miniCPAP Begin by unpacking all items from the Transcend travel bag and inspect them to ensure they were not damaged during shipment. Report any missing or damaged items to the home healthcare provider that provided the product to you.
  • Page 13: What's Not Included (All Sold Separately)

    Transcend 3 miniCPAP User Manual Page 13 What’s not included (all sold separately) • Transcend Mobile Power Adapter • Transcend P Battery • Transcend P Battery • Transcend Portable Solar Battery Charger (to be used as an alternative charging source for Transcend P...
  • Page 14: Description Of Transcend 3 Minicpap Components

    Description of Transcend 3 miniCPAP components Transcend 3 miniCPAP Device The Transcend 3 miniCPAP comes ready to generate and regulate continuous positive airway pressure therapy for delivery to the interface (mask). An external power source connects to the Transcend 3 miniCPAP to supply power to the device.
  • Page 15: Power Connection Jack And Usb Port

    DC power source to operate the Transcend 3 miniCPAP. A variety of power sources may be used to power the Transcend 3 miniCPAP. An AC to DC converting power supply is provided with your device and should be used when powering the device by line (wall outlet) power.
  • Page 16: Air Inlet Filter

    Transcend 3 miniCPAP User Manual Page 16 Air Inlet Filter During therapy operation ambient air is drawn into the Transcend 3 miniCPAP through an Air Inlet Filter. The Filter Assembly should be replaced minimally after 6 months of use. Replace more frequently as desired.
  • Page 17: Assembling The Transcend 3 Minicpap

    Page 17 Assembling the Transcend 3 miniCPAP Attach the air supply tube to the air outlet on the Transcend 3 miniCPAP device. Connect the mask to the opposite end of the air supply tube. Plug the power supply barrel connector into the Transcend 3 miniCPAP power jack on the rear of the device.
  • Page 18: Powering The Transcend 3 Minicpap

    Transcend 3 miniCPAP User Manual Page 18 Powering the Transcend 3 miniCPAP There are three choices for powering your Transcend 3 miniCPAP device: • Using the Multi-Plug Universal Power Supply (PSA2) • Using the optional Transcend (P or P Battery •...
  • Page 19: Using The Multi-Plug Universal Power Supply

    The Transcend 3 miniCPAP power-up LED flash sequence should initiate. Once the power-up LED flash sequence is complete the LED lights will turn off. This sequence indicates that power is being supplied to the Transcend 3 miniCPAP and that it has successfully entered Standby Mode.
  • Page 20: Using The Transcend P Or P Battery

    P battery outlet cable to the Transcend 3 miniCPAP by inserting the barrel connector into the power jack on the Transcend 3 miniCPAP. Be sure the power-up LED flash sequence completes indicating that power is being supplied to the Transcend 3 miniCPAP and that it has successfully entered Standby Mode.
  • Page 21: Explanation Of The Transcend P And P Battery Led Lights

    Charge the battery fully before the first use. Do not connect the Battery to Transcend 3 miniCPAP during the initial charge. Note To maintain maximum battery performance Somnetics recommends using the Battery in-line with the AC Power Supply or Mobile Power Adapter during therapy even if the battery is fully charged.
  • Page 22 Transcend 3 miniCPAP User Manual Page 22 • Connect the power outlet barrel connector from the Universal Power Supply or Mobile Power Adapter to the power connection jack on the P or P Battery. • Connect the Universal Power Supply or Mobile Power Adapter to a power source.
  • Page 23: Using The Transcend Mobile Power Adapter (Mpa1)

    Make certain that the cables are securely connected to the Mobile Power Adapter and to the power jack on the back of the Transcend 3 miniCPAP. It may be necessary to unplug the cables and reconnect to ensure a good connection.
  • Page 24: Using The Transcend 3 Minicpap

    The Transcend 3 miniCPAP operational status is displayed by LED illumination states. When a power source is connected to the device the Transcend 3 miniCPAP power-up LED flash sequence should initiate. Once the power-up LED flash sequence is complete the LED lights will turn off.
  • Page 25: Starting Therapy

    Ramp Mode until it reaches the prescribed therapy pressure. Be sure the Transcend 3 miniCPAP is in On Mode. If not, press the power button. Adjust your mask to eliminate mask leaks. Press the Ramp button. The pressure will drop to the Ramp starting pressure and will gradually increase over a preset length of time until reaching the prescribed therapy pressure.
  • Page 26: Using The Ezex Function

    Transcend 3 miniCPAP User Manual Page 26 Note In the event of power loss during ramp, the Transcend 3 miniCPAP will resume at the full prescribed pressure as soon as power is restored. Using the EZEX function The EZEX function is a special feature that decreases therapy pressure on exhalation.
  • Page 27: Replacing The Filter Media

    Page 27 Replacing the Filter Media The Transcend 3 miniCPAP Filter Media should be replaced every six months at minimum. Remove the filter assembly by pulling the tab on the back of the device to release it. Then pull the filter assembly away from the device.
  • Page 28: Caring For Your Transcend 3 Minicpap And Components

    Submerge a lint-free cotton cloth into the detergent solution. Wring excess water from the cloth then wipe the exterior of the Transcend 3 miniCPAP device for approximately 20 seconds using a gentle, back and forth wiping motion from the front to back of the device.
  • Page 29: Cleaning Of Accessories

    Transcend 3 miniCPAP User Manual Page 29 Exterior cleaning of the device should be performed as follows: Product Periodic Cleaning Cycle Product Service Life Transcend 3 miniCPAP 2x/Month 5-Year Cleaning of Accessories The following accessory should be cleaned with a 5% solution of mild liquid detergent in distilled water (1.6 fl oz liquid cleaning detergent per quart of distilled water).
  • Page 30 Transcend 3 miniCPAP User Manual Page 30 Apply a new bacterial/viral filter to the air outlet of the Transcend 3 miniCPAP and attach a new air supply tube before providing the device to a new user. The air supply tubing, mask assembly and bacterial/viral filters should...
  • Page 31: Software User Guide

    Page 31 Software User Guide _______________ Your Transcend 3 miniCPAP device comes with easy-to-use software that helps you adjust settings, observe, print and send reports and store all your data so you can access it any time. Simple Steps and you’re ready to use your software.
  • Page 32: Connecting Your Transcend 3 Minicpap

    Page 32 STEP Connect your Transcend 3 miniCPAP device to a power source, then connect it to your computer using the supplied mini-USB cord. The mini end of the USB cord goes into the opening next to the power cord on the device; open the rubber cover to access the opening.
  • Page 33: Configuration For Device Settings

    Transcend 3 miniCPAP User Manual Page 33 Now You’re Ready to Use the Transcend 3 miniCPAP Software You can see on the Home screen interface that there are two primary areas to choose from: Configuration or Compliance Manager. Let’s explore each one.
  • Page 34: Compliance Manager

    This setting will show either Fixed or Auto setting, depending on which device you have (Auto mode only available on Transcend 3 miniCPAP Auto device). If you have a fixed pressure device, one therapy pressure is available for adjustment. If setting an Auto device, the ‘Auto’...
  • Page 35: General Functions

    Transcend 3 miniCPAP User Manual Page 35 You can see the primary data summaries are: ○ Usage ○ Leak Summary ○ Pressure Summary ○ AHI-Events per Hours of Use ○ Patient Therapy Settings General Functions Filtering by Date Tailor the date range of information by clicking on the calendar icons next to choose the desired start and end dates.
  • Page 36: Generate Compliance

    Transcend 3 miniCPAP User Manual Page 36 Generate Compliance (Last 30 days of compliance report) The 30 Day Compliance button will automatically pull the last 30 day window of compliance data that has at least 70% compliance (4+ hours of therapy 70% of the time).
  • Page 37: Day List

    Transcend 3 miniCPAP User Manual Page 37 Choose a View At the top of the page you will also see three grey tabs: Daily List, Trend Graph and Day Graph. Daily List When you select the Daily List tab you will see a general summary for the date range you have indicated.
  • Page 38: Trend Graph

    Transcend 3 miniCPAP User Manual Page 38 Trend Graph When you click the Trend Graph tab at the top of the page, you’ll see the data in graphic form. Simply click on any bar graph and you will automatically see trend ranges by month, week or day each time you click (this change in date range is reflected in the Trend Range section).
  • Page 39: Day Graph

    Transcend 3 miniCPAP User Manual Page 39 Day Graph Select the Day Graph tab is used to show data that has been drilled down to the day in the Trend Graph tab or you can simply go to this tab to select a particular day’s worth of data. You can choose the day you would like to view by selecting the date in the Selected Date field.
  • Page 40: Fault And Alert Codes

    Fault and alert codes Fault codes When the Transcend 3 miniCPAP encounters a fault the processor resets and enters a fault loop. In this loop, the device repeatedly flashes the yellow fault LED to indicate the specific fault encountered. If the device is reset or power cycled while in fault mode, it re- enters fault mode upon power up.
  • Page 41: Alert Codes

    Transcend 3 miniCPAP User Manual Page 41 Device Fault Error Comments Flashes Low power While delivering therapy, the device determined that the 12 times blower was stalled (see above). However, a check of the voltage indicates there may not be sufficient power to spin the blower.
  • Page 42: Troubleshooting

    Transcend 3 miniCPAP User Manual Page 42 Troubleshooting Problem Probable Cause Solution Discomfort due to a feeling Device pressure may be set too Breathe slowly through your nose with of high pressure. high. your mouth closed. Use the ramp pressure, if available.
  • Page 43: Appendix: Part Numbers

    Transcend 3 miniCPAP User Manual Page 43 Appendix: Part numbers This section presents three topics: • Disposable parts • Accessories • Replacement parts Disposable parts Item Part number Item Part number Transcend 3 Filter Assembly 503109 Standard 6-foot hose 503081...
  • Page 44: Appendix: Specifications

    Transcend 3 miniCPAP User Manual Page 44 Appendix: Specifications This section presents the following topics: • Transcend 3 miniCPAP • AC power supply - PSA2 • Mobile power adapter (optional) - MPA1 • Batteries • Transcend 3 miniCPAP performance •...
  • Page 45: Transcend 3 Minicpap Performance

    Transcend 3 miniCPAP User Manual Page 45 Transcend 3 miniCPAP performance Working pressure range: 4 to 20 cm H Accuracy of pressure setting: ±1 cm H O or ±10%, whichever is greater Maximum system shutdown pressure: 30 cm H Ramp time duration:...
  • Page 46: Manufacturer's Declaration

    • IEC 60601-1 Compliance Electromagnetic emissions The Transcend 3 miniCPAP is intended for use in the electromagnetic environment specified below. The customer or the user of the system should ensure that it is used in such an environment. Emissions test Compliance Electromagnetic environment—guidance...
  • Page 47 Transcend 3 miniCPAP User Manual Page 47 The Transcend 3 IEC 60601 test Compliance level Electromagnetic miniCPAP is level environment—guidance intended for use in the electromagnetic environment specified below. The customer or the user of the system should ensure that it is used in such an environment.
  • Page 48: Iec 60601-1 Compliance

    Transcend 3 miniCPAP User Manual Page 48 IEC 60601-1 compliance Protection against electric shock: Class II Type BF Degree of protection against ingress of IP22. Protected against ingress of solid foreign objects water: greater than or equal to 12.5 mm in diameter. Vertically falling drops shall have no harmful effects.
  • Page 49: Appendix: Limited Warranty

    Transcend 3 miniCPAP User Manual Page 49 Appendix: Limited warranty Somnetics warrants its products to be free of defects in materials and workmanship and will perform in accordance with the product specifications for a period specified in the following table:...
  • Page 50 Transcend 3 miniCPAP User Manual Page 50 Somnetics International, Inc. 33 5th Avenue NW, Suite 500 New Brighton, Minnesota 55112 USA http://www.mytranscend.com/ Phone: 651.621.1800 Toll-free: 877.621.9626 Fax: 651.204.0064 103940 Rev B...