The control panel of the Transcend Auto has two pushbuttons that activate the blower and
initiate the pressure ramp feature. There are also two LED lights, including a green LED
for indicating normal operational modes and a yellow LED for indicating fault conditions.
The Transcend Auto operational status is displayed by LED illumination states.
When a power source is connected to the device the Transcend Auto power-up LED flash
sequence should initiate. Once the power-up LED flash sequence is complete the LED
lights will turn off. This sequence indicates that power is being supplied to the Transcend
Auto and that it has successfully entered Standby Mode. During therapy delivery, the LED
lights remain off to avoid disturbing the patient and/or bed partner.
Note
If the Transcend Auto loses power while delivering therapy it will resume delivering
therapy as soon as power is restored and you press the power button. The device
will repeat the power-up LED flash sequence prior to the blower restarting.
Standard user modes
Normal operation consists of four modes:
Off
When the device is not connected to a power source the device is off. Control
panel LEDs are both off.
Standby
When power is applied to the device it completes the power-up LED sequence
and enters Standby Mode. Standby Mode is also initiated by pressing the
power button when the device is in On Mode or if the mask is removed while in
On Mode. As long as power is supplied to the device it will remain in Standby
Mode until On or Drying Mode is initiated.
On
When in On Mode the blower is working and regulated device therapy pressure
is being generated. The LED lights remain off. On Mode is initiated by pressing
the power button when the device is in Standby Mode and the mask is worn by
the patient.
Drying
To initiate Drying Mode depress the Ramp button and press the Power button
simultaneously. When in Drying Mode the blower runs at a low speed for 30
minutes. During Drying Mode the LED lights remain off and blower pressure is
not regulated to provide therapy.