Transcend Auto User Manual
Device
Fault
LED
LED
Off
Flashes
13
times*
Off
Flashes
14
times*
*
If this fault occurs, check all tubes and hoses to ensure that they are properly connected. If the fault
occurred during therapy, make sure the mask is properly fitted and not leaking. Check that all filters are
properly installed and not excessively dirty.
Alert codes
Alerts are similar to faults but they do not reset the processor when they occur (i.e., the Auto
device continues to deliver therapy pressure).
Device
Fault
LED
LED
Off
Flashes
15 times
Error
Comments
Processor over-
The processor on-chip temperature sensor has reported an
temp
excessive temperature.
Blower over-
The blower thermistor has reported an excessive temperature.
temp
Error
Comments
Cannot regulate
Device repeatedly detected the monitored pressure outside of
the therapy target range. The device attempts to continue to
deliver therapy. The alert is an indication that filters and the
mask should be inspected for cleanliness and proper fit.