Cleaning The Filter Media And Filter Frame - Somnetics Transcend Auto User Manual

Sleep apnea therapy system
Hide thumbs Also See for Transcend Auto:
Table of Contents

Advertisement

Transcend Auto User Manual
Rinse the air supply tube in distilled water by fully immersing. Lift, then lower, the
5.
ends of the tube while the tube is filled with water. Repeat this motion for
approximately 10 seconds to remove cleaning agent residue.
Dry the accessory by wiping with dry, lint-free cotton cloth
6.
dry as needed.
Dry the outside of the air delivery tube with a dry, lint-free cotton cloth. Allow the tube
7.
to air dry until the inside of the tube is dry. Length of drying time will depend on
ambient conditions.
Note:
It is recommended that fabrics air dry at room temperature for 30-60 minutes prior
to reassembly or use.

Cleaning the Filter Media and Filter Frame

Clean the foam Filter Media and Filter Frame at least once per week. Follow these steps to
clean:
Begin by removing the Filter Media and Filter Frame from the device (see
1.
the Filter Media
Mix a solution of 5% mild liquid detergent in distilled water (1.6 fl oz liquid detergent
2.
per quart of distilled water). Mild detergent should contain biodegradable anionic
surfactants and no phosphate.
Fully immerse the
3.
While immersed, thoroughly wipe each surface of the Filter Frame with a lint-free
4.
cotton cloth. Apply firm pressure and ensure contact with all accessible contact
surfaces to adequately remove soil buildup.
Rinse the Filter Frame by immersing in distilled water and gently moving it forwards
5.
and backwards for approximately 10 seconds.
Use a clean, dry, lint-free cotton cloth to dry all surfaces of the Filter Frame
6.
Filter Media
To clean the
7.
.
solution
Wipe the Filter Media with cloth to remove any visible residue. Wipe all surfaces for a
8.
minimum of 10 seconds.
Rinse the cotton cloth in distilled water to remove residual cleaning solution.
9.
Wring excess water from the cloth then wipe all surfaces of the Filter Media to remove
10.
any cleaning solution remaining on its surface.
Dry surfaces by wiping with a dry, lint-free cotton cloth.
11.
Allow the Filter Media to air dry at room temperature for approximately 30-60 minutes
12.
before reassembly. Drying time will depend on ambient conditions.
for instructions on filter removal and replacement).
Filter Frame
in the detergent solution.
, first submerge a lint-free cotton cloth into the cleaning
.
Allow the accessory to air
Replacing
.
Page 28