Displayed Messages - Devatec ElectroVap ELMC 1 SM CYL Technical Manual

Electrode steam humidifier electrovap mc2 series
Hide thumbs Also See for ElectroVap ELMC 1 SM CYL:
Table of Contents

Advertisement

Maintenance - Service & alerts
1
2
INSPECTION DUE
50 H − SEE MANUAL
SERVICE DUE
SEE MANUAL
SERVICE OVERDUE
SYSTEM OFF
REPLACE CONTACTORS
10000 HR RUN
E
LECTRO
Press Select button 1 to rotate between the menus and
press the up or down buttons to enter the desired menu.
3
50 HOURS AFTER SWITCHING ON
- A « 1st inspection see after manuel » warning message is shown on the
window.
- The unit works on.
- The remote maintenance contact is trigged on.
- The remote general failure contact is off.
- To reset the warning message, press key 3 for over 5 seconds
- This 50 hours timer cannot be suppressed nor modified.
SERVICE TIME ELAPSED (service message)
- A « CYLINDER MAINTENANCE - SEE AFTER TECH MANUAL » service mes-
sage is shown.
- The unit works on.
- The remote maintenance contact is trigged on.
- The remote general failure contact is off.
Maintenance to the cylinder(s) (s.a. page 54) and to the valves (s.a. pages
55-56) should be done.
- This service message is removed only when the manual draining is completed
(about 6 mn). The « DRAINING CYCLE OVER » service message is then dis-
played.
- This timer (300 H factory preset) can be adjusted (access code 2.3.4
« MAINTENANCE FREQUENCY (HR) » but not removed.
SERVICE TIME OVERDUE (service message)
- If the previous alert has not been cleared, a new alert « SERVICE OVERDUE /
SYSTEM OFF » is shown 100 hours thereafter.
- The humidifier is brought to a halt.
-The remote maintenance contact is trigged on.
-The remote general fault alert is trigged on.
Maintenance to the cylinder(s) (s.a. page 54) and to the valves (s.a. pages
55-56) must be done.
- This alert message is removed only when the manual draining is completed (in
about 6 mn). A « DRAINING CYCLE OVER » message is then shown on the
display. Press button 1 to bring the humidifier back to operation.
- This 100 H timer cannot be suppressed nor modified.
CONTACTOR TO REPLACE (service message)
- «10000 H REPLACE CONTACTOR» is shown after he humidifier has been
running for 10000 hours. This means that the contactor(s) has/have been
trigged on and off for 10000 hours and that its/their replacement is/are highly
recommended.
- The unit works on.
- The remote maintenance contact is trigged on.
- The remote general fault contact is trigged on.
- To reset this alert message, press key 3 for over 5 seconds
- This 10000 H timer cannot be suppressed nor modified.
V
MC2
AP
51
Pictures for illustrating purposes only
:

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents