Set/Calibrate Flow Rate
• Calibrate the fl ow rate with a representative sampling medium in line.
• Allow pump to equilibrate after moving it from one temperature extreme to another.
1.
Ensure the pump has run for 5 minutes before calibrating. Using 3/8-inch ID fl exible tubing and appropriate adapters, connect the
inlet of the pump to the outlet of a representative sampling medium; connect a calibrator to the inlet of the representative sampling
medium. See the example of a calibration train above.
2.
Scroll to any run time preset. Press and hold the CAL butt on for 2 seconds. The pump will start running and
a fl ashing CAL will appear on the LED.
Press  or  to increase or decrease the fl ow until the desired fl ow rate is displayed on the calibrator.
3.
4.
Press
. A fl ashing Stor will appear on the LED.
• To save new sett ing: Press [
• To ignore new sett ing: Press
Note: A security code is not needed to change fl ow rate. It is only required when changing factory sett ings. To change
factory sett ings, see Advanced Operation.
Tubing
Calibrator outlet
Calibration train with a spore trap cassette
]. The pump will stop running.
. The pump will stop running.
www.skcinc.com
Note: Use a second cassette adapter to
attach calibrator to a spore trap cassette.
Spore trap cassette
Cassette adapter
Pump inlet
Digital display:
Calibration mode
Digital display:
Store setting
Form 40079 Rev 1705