Getting Started
Charge the Battery
1.
For a complete charge, ensure the pump is not running. Insert the charger plug into
the charging jack on the pump.
2.
Insert the charger into a wall outlet. A red LED will fl ash on the pump display to
indicate the unit is charging. When charging is complete, the LED will stop fl ashing
and the pump will go to sleep. The batt ery charges completely in 5 hours. Note: The
QuickTake 30 can be operated using AC power. See Operate from AC Power.
Determining Battery Charge
Batt ery Status Indicators
Full charge; approximately 75 to 100% batt ery capacity remaining
Batt ery is charged enough to operate the pump; approximately 25 to 75% batt ery capacity remaining.
Batt ery charge is low (charge batt ery); approximately 1 to 25% batt ery capacity remaining.
Low Batt ery Fault. Pump will stop running, beep, and go to sleep in 10 seconds.
Notes and Cautions
• After charging the battery pack, it is good practice to run the pump for approximately 5 minutes before calibrating. This ensures
the battery is in a more steady-state condition and improves the agreement in pre and post-sampling calibrations.
• The pump will not sleep during charging. Connecting a sleeping pump to the charger will wake it up from sleep.
• The AC charger/adapter can be used to extend battery run time, but it is not a battery eliminator. Therefore, it will not provide
indefi nite run times.
• Do not operate or charge the pump in hazardous locations.
• Use only the SKC-approved charger for this pump. Use of an unapproved charger may damage the battery and the pump
and voids any warranty.
• The battery pack may be kept on the SKC-approved charger for an indefi nite time.
• Ensure proper orientation of the charging cable before plugging it into the charging jack. Improper orientation/contact will
short circuit the battery and voids any warranty.
• Short-circuiting the battery pack will render it immediately inoperative.
• Failure to follow warnings and cautions voids any warranty.
See www.skcinc.com/instructions/1918.pdf for more information on SKC pump lithium-ion batteries.