Smeg WD2145 User Manual page 62

Instrument washers
Hide thumbs Also See for WD2145:
Table of Contents

Advertisement

ALARM ID
DESCRIPTION
A-:68
P1 jerry can empty
A-:69
P2 jerry can empty
A-:70
P3 jerry can empty
Internal memory data storage
AF:73
error
Water leak from chamber. Alarm
AF:74
only enabled with the Aquastop
accessory fitted.
No salt in softener. This alarm
does not appear on the Display but
AF:75
is recorded in the appliance
memory; the relative LED
illuminates on the display.
Intake water temperature over
45°C; prewash temperature must
AF:77
be below 45°C. (Alarm generally
disabled on GW series products.)
Restore fail. Problem of
AF:78
motherboard.
AF:79
Program not compatible.
Dryer temperature reading higher
AF:84
than actual temperature. (only for
WD4060, WD4190 series)
AF:91
Internal memory full.
User Manual
Check that there is detergent in the jerry can or that the level
sensor is operating correctly.
This is a warning and not an alarm as such; the washing cycle can
be started with this message present (i.e. overriding the warning)
by pressing the Start/Stop button again (hold it down for 2
seconds).
Activated at the end of the cycle and when the user attempts to
start a new program. If two or more similar alarms (e.g. cans P1
and P2 empty), the Start/Stop button will have to be pressed
twice or more before the cycle can be started. (The device's
internal memory records the event but allows execution of the
cycle).
As for "A-:68"
As for "A-:68"
For the user: follow the DEFAULT PROCEDURE described above.
If the alarm does not disappear, follow the RESET PROCEDURE.
Turn off the water supply taps.
Contact the After-Sales Service.
Add salt to the softener salt tank inside the chamber, accessible
with the door open.
The alarm occurs when the temperature is over 45°C at the start
of the cycle.
Wait for the appliance to cool before starting a new cycle.
For the user: follow the DEFAULT PROCEDURE described above.
If the alarm does not disappear, follow the RESET PROCEDURE.
This alarm is triggered if a program is created using the
TRACELOG software using procedures not compatible for the
proper performance of the cycle. Reconsider the program
created phase by phase; it may be useful to compare it with an
original program provided as reference (see attached
PROGRAMS TABLE doc.).
Pay special attention to the version of the appliance being used;
e.g. do not activate peristaltic pump P3 when writing the
program if this pump is not installed.
Wait for the alarm to be resolved automatically.
During automatic management of the AF:84 alarm the RESET
cycle cannot be started.
Do not disconnect the electricity supply to the device: the
overheating problem is being dealt with automatically, with a
heating element cooling cycle.
The DEFAULT PROCEDURE described above cannot be
implemented until the end of the automatic process.
This alarm only occurs if the appliance parameter that prevents
overwriting in the memory is enabled.
With the default setting, this alarm will not occur.
Memory space must be cleared to allow use of the appliance to
continue: this can be done by connecting to the appliance by
means of the RS-232 serial port and using the TRACELOG
software.
WD2145-WD1160-WD4060-WD4190
USER ACTION
Page 62 - 85