Set PARAMETER
Power Supply
The parameter indicates whether the connection is:
- single-phase: parameter setting "1"
- three-phase: parameter setting "3"
-A
N.B.: the value "3" refers to both 400V 3N~ / 50 Hz and 230V
3~ / 50Hz connections.
Warning: this parameter is only used for the correct setting of
times remaining.
Memory Data Overwriting
- Overwriting enabled: parameter setting "1"
- Overwriting disabled: parameter setting "0"
The parameter refers to whether or not the device's internal
memory, which keeps a record of all the cycles performed and
-b
the alarms which have occurred, can be overwritten.
WARNING
If overwriting is disabled, alarm AF:91 is triggered once the
memory is full. In this case the appliance will not perform any
more cycles until the data from the memory are downloaded;
this can only be done using the TRACELOG software.
Superuser dosage setting disabling parameter.
(Parameter introduced from Main Firmware version
4.6.xx.yy).
If "-d: 0" the superuser is able to modify the dosages
-d
If "-d: 1" the superuser is not able to modify the dosages but
can display them.
Default value: "-d: 0".
This Parameter can only be modified by the authorised
engineer.
Parameter for setting presence or absence of the WD-LANE
accessory.
(Parameter introduced from Main Firmware version
4.6.xx.yy).
-E
Sets the serial port communication speed
"-E: Pr" = 9600 bps (for Printer connection)
"-E: Ln" = 11500 bps (for connection to WD-LANE - LAN
accessory).
Test cycle
(Parameter Introduced from the Main Firmware version
1.11.0.33) The parameter can be used by authorized
technicians only; it allows the display and the start of a Test
program, on the display "Pr t". This test program performs a
short washing cycle and activates all the loads of the device.
The cycle is defined to be used for machine verification after
-t
components replacement.
- Parameter value "--", you do not have access to the cycle "Pr
t"
- Parameter Value "on", the technician gets access to the Test
Program "Pr t".
In order not to run the Test cycle "Pr t": switch Off the device
to reset its condition..
In the following pictures some examples of "Set" parameters are featured
User Manual
DESCRIPTION
WD2145-WD1160-WD4060-WD4190
POSSIBLE SUPERUSER ACTION
With set
With set
parameter
parameter
"-d=0"
"-d=1"
Display Only
Display Only
Display and
Display and
Modification
Modification
Display Only
Display Only
Display and
Display and
Modification
Modification
Display Only
Display Only
Page 43 - 85

Hide quick links:

Advertisement

Table of Contents
loading

This manual is also suitable for:

Wd1160Wd4060Wd4190

Table of Contents