Smeg WD2145 User Manual page 46

Instrument washers
Hide thumbs Also See for WD2145:
Table of Contents

Advertisement

================================================================================
2012/10/17 10:39:46 START PROGRAM: 7
Cycle N.: 288
2012/10/17 10:40:17 START PHASE N.: 1
2012/10/17 10:41:12 Inflow Detergents
2012/10/17 10:42:26
2012/10/17 10:42:26 Start heating
2012/10/17 10:56:12
2012/10/17 10:57:36
2012/10/17 10:57:52
2012/10/17 10:59:43
2012/10/17 10:59:58
2012/10/17 11:01:53
2012/10/17 11:02:08
2012/10/17 11:04:12
2012/10/17 11:04:26
2012/10/17 11:05:53 Ao>=15983
2012/10/17 11:05:53 Start heating
2012/10/17 11:05:53
2012/10/17 11:07:49 START PHASE N.: 2
2012/10/17 11:08:41 Inflow Detergents
2012/10/17 11:09:34
2012/10/17 11:09:34 Start heating
2012/10/17 11:11:34
2012/10/17 11:11:34 Start heating
2012/10/17 11:11:34
2012/10/17 11:12:33 START PHASE N.: 3
2012/10/17 11:13:19 Inflow Detergents
2012/10/17 11:13:41
2012/10/17 11:13:41 Start heating
2012/10/17 11:15:41
2012/10/17 11:15:41 Start heating
2012/10/17 11:15:41
2012/10/17 11:16:41 START PHASE N.: 4
2012/10/17 11:17:26 Inflow Detergents
2012/10/17 11:17:49
2012/10/17 11:17:49 Start heating
2012/10/17 11:25:47
2012/10/17 11:26:30 Start heating
2012/10/17 11:26:30
2012/10/17 11:27:29
2012/10/17 12:04:47
2012/10/17 12:09:07 END PROGRAM: 7
================================================================================
(The first phase is referred to as "PHASE N.1" with effect from main firmware version 2.0.0 of 23/09/2013,
User Manual
Oper: 255
P1:0.40%
TL1: 22.7
TCL: 22.5
Target temp. 93
TL1: 94.1
TCL: 93.6
TL1: 93.9
TCL: 93.4
TL1: 94.1
TCL: 93.6
TL1: 93.9
TCL: 93.4
TL1: 94.1
TCL: 93.6
TL1: 93.9
TCL: 93.4
TL1: 94.1
TCL: 93.6
TL1: 93.9
TCL: 93.4
TL1: 94.1
TCL: 93.5
Target temp. 0
TL1: 94.1
TCL: 93.6
P2:0.20%
TL1: 43.3
TCL: 42.3
Target temp. 0
TL1: 59.7
TCL: 59.2
Target temp. 0
TL1: 59.7
TCL: 59.2
TL1: 36.9
TCL: 36.5
Target temp. 0
TL1: 49.9
TCL: 49.5
Target temp. 0
TL1: 49.9
TCL: 49.5
TL1: 33.3
TCL: 32.9
Target temp. 75
TL1: 76.2
TCL: 75.7
Target temp. 0
TL1: 76.9
TCL: 76.5
Start drying: TA_TARGET: 120 TA1: 73
End drying: TA1: 89
fig. 35 – Example of cycle data stored in the appliance's memory.
it was formerly referred to as "PHASE N.0")
WD2145-WD1160-WD4060-WD4190
The double line marks the
start and end of a program
Consecutive number of cycle
First phase, called "1"
P1 detergent dispensing
Phase Target temperature
A0 value reached
Second phase, called "2"
The two tank temperature
probes are called:
TL1 – working probe
TCL – control probe
Drying phase, with Target
temperature in Dryer duct