Source Current Mode Programming And Readback Accuracy; Table 2-5 Source Current Mode Programming And Readback - Keysight U2722A Service Manual

Usb modular source measure units
Hide thumbs Also See for U2722A:
Table of Contents

Advertisement

Source current mode programming and readback accuracy

Table 2-5
Source current mode programming and readback accuracy
Iteration
Test point
1
0.5 μA
2
3
5 μA
4
5
50 μA
6
7
0.5 mA
8
9
5 mA
10
11
60 mA
12
Keysight U2722A/U2723A Service Guide
Clamp
Current range
vol tage
20 V
1 μA
2 V
20 V
10 μA
2 V
20 V
100 μA
2 V
20 V
1 mA
2 V
20 V
10 mA
2 V
20 V
120 mA
2 V
Performance Verification and Adjustment
I
– I
Vol tage range
meas
20 V
2 V
20 V
2 V
20 V
2 V
20 V
2 V
20 V
2 V
20 V
2 V
I
– I
drive
rdbk
meas
±1.275 nA
±1.275 nA
±12.75 nA
±12.75 nA
±112.5 nA
±112.5 nA
±1.125 μA
±1.125 μA
±11.25 μA
±11.25 μA
±0.16 mA
±0.16 mA
2
Limit
33

Advertisement

Table of Contents
loading

This manual is also suitable for:

U2723a

Table of Contents