7.2.

Optional Extra: Handheld Enclosure

The handheld FRP enclosure is designed for fixed or secure monitoring. The inlet nozzles are
specifically designed to eliminate dust and reduce water ingress whilst minimising destruction of part
per billion gas concentrations or measurement variations due to outgas contamination from the
enclosure. The sensor heads are specifically calibrated to take into account the longer flow path.
Sensor heads fitted to this product must be the Type 2 sensor heads that are capable of operating
with the longer flow path.

7.2.1. Specifications

Dimensions
Height (including cable gland)
Width
Depth
Materials
Mounting
Operating temperature
Inlet
Temp/RH sensor
connection

7.2.2. Replacing the Sensor Head

1. Undo the four lid screws at each corner using a flat headed screw driver and remove the lid.
Kanomax Gasmaster User Guide
MRK-D-0002 V3
283 mm
180 mm
90 mm
Fibre-glass reinforced polycarbonate base
Polycarbonate lid
Screw fixture
-35°C to 120°C
Power Connection
Outlet
Data logging
connection