3. Verify that the AC power cable is not damaged, and that the power-source outlet provides a
protective earth contact.
CAUTION
Always use the three-prong AC power cord supplied with this product. Failure to
ensure adequate earth grounding by not using this cord may cause personal injury or
product damage.

STEP 3. Install the Analyzer

IMPORTANT Please review all safety information located in Chapter 4 , "Safety, Regulatory, and
Environmental Information," before proceeding with installing and using the analyzer.
The analyzer can be installed on a bench top or in an equipment rack. When installing an analyzer
equipped with handles in an equipment rack, use Option 1CP. When installing an analyzer not
equipped with handles in an equipment rack, use Option 1CM.
In all installations, consider the following ventilation requirements when deciding where to set up your
analyzer.
CAUTION
Ventilation Requirements: When installing the product in a cabinet, the convection
into and out of the product must not be restricted. The ambient temperature (outside
the cabinet) must be less than the maximum operating temperature of the instrument
by 4°C for every 100 watts dissipated in the cabinet. If the total power dissipated in the
cabinet is greater than 800 watts, then forced convection must be used.
Refer to the section below that applies to your installation.
To Install on a Bench Top
When you install the analyzer for bench top use, you may wish to attach the supplied handles to the
analyzer. If you wish to install the handles, follow the instructions provided with the handle kit.