Installing Your Analyzer

STEP 1. Check the Shipment

Unpack the contents of the shipping container. Keep the packaging material in case the
1.
analyzer needs to be returned to the factory.
WARNING
The network analyzer is heavy. It is recommended that two individuals, or a
mechanical lift, be used to lift or transport the instrument.
NOTE
The PNA is shipped with ESD-safe protective end caps on the test ports and metal covers for
the front-panel jumpers. Save these items for reuse to ensure shipment safety of the PNA.
2. Carefully inspect the analyzer to make sure that it was not damaged during shipment. If
damaged, refer to "Contacting Agilent" on page 15.
3. Use the Contents List in the shipping container to verify the completeness of your shipment. If
not complete, refer to "Contacting Agilent" on page 15

STEP 2. Meet Requirements

Before installing the PNA on a bench top or in an equipment rack, first make sure that the electrical
and environmental requirements of the location are met.
CAUTION
This instrument has autoranging line voltage input. Be sure the supply voltage is within
the specified range.
1. Ensure the available AC power source meets the following requirements:
NOTE
The instruments can operate with mains supply voltage fluctuations up to ±10% of the
nominal voltage.
For E836xC, N5230C, and N523x models:
Voltage: 100/120/220/240 VAC
Frequency: 50/60/400 Hz for 100/120 VAC; 50/60 Hz for 220/240 VAC
Power (maximum consumed): 350 W
For all other models:
Voltage: 100/120/220/240 VAC
Frequency: 50/60/400 Hz for 100/120 VAC; 50/60 Hz for 220/240 VAC
Power (maximum consumed): 450 W
2. Ensure the operating environment meets the requirements found in "Environmental
Information" on page 20.
NOTE
Some analyzer performance parameters are specified for 25°C. Refer to the Help system in
the analyzer for the complete specifications.