Environmental Information

Samples of this product have been type-tested in accordance with the Agilent Environmental
NOTE
Test Manual and verified to be robust against the environmental stresses of storage,
transportation and end-use; those stresses include but are not limited to temperature,
humidity, shock, vibration, altitude and power-line conditions. Test methods are aligned with
IEC 60068-2 and levels are similar to MIL-PRF-28800F Class 3.
Parameter
Temperature
Operating
Storage
Error-corrected range
Operating Altitude
Relative humidity

Warranty

The actual warranty on your instrument depends on the date it was ordered as well as whether or not
any warranty options were purchased at that time. To determine the exact warranty on your
instrument, contact Agilent Technologies with the model and serial number of your instrument. Refer
to "Contacting Agilent" on page 15
Required Values/Ranges
For Indoor Use ONLY
0 °C to +40 °C
The instrument powers-up and displays no error messages within this temperature
range (except for "source unleveled" error message that may occur at temperatures
outside the specified performance temperature range of 25
−40 °C to +70 °C
23 °C ±3 °C with less than 1 °C deviation from calibration temperature.
0 to 4,600 meters (≈15,000 feet)
Type tested, 0% to 95% at 40 °C, non-condensing
In conditions of very high humidity (below 95%) at changing temperature,
there is a small risk of internal condensation that could cause the CPU
real-time clock to temporarily stop incrementing time.
21
± 5 °C).