Performance Tests
Instrument Accuracy Test
Test Setup
Procedure
The following procedure should be performed for the Instrument Accuracy
Test.
Note
2-8
Power Meter
CHANNEL A
Figure 2-2: Instrument Accuracy Test Setup
The procedure details the key presses required on the
Agilent E4418B. For the Agilent E4419B the equivalent key
presses should be performed on both channels.
1. Connect the equipment as shown in Figure 2-2.
2. Switch the power meter on.
Preset
3. Press
then
Local
dBm/W
4. Press
,
W
5. Set the range calibrator as follows:
RANGE............................................ 3 µW
POLARITY ...................................... NORMAL
FUNCTION..................................... STANDBY
LINE ............................................... ON
Agilent 11730A
.
Confirm
for a reading in watts.
Agilent E4418B/E4419B Service Guide
Range
Calibrator
POWER
METER