Agilent Technologies E4418B Service Manual page 144

Power meters
Hide thumbs Also See for E4418B:
Table of Contents

Advertisement

Measurement Uncertainty Analysis – Power Reference Level Test
Measurement Introduction
Consider the model below:
An estimate of the moduli of Pd and Ps are each less than a specified value. Ps and Pd
each lie within a circle of radius P. Assuming Pd and Ps have equal probability of
lying anywhere within the circle the standard uncertainty of Mismatch becomes
(
u Ms
Ps is the reflection coefficient of the source, which is specified as less than 1.05.
The reference is designed to have a VSWR of less than 1.05.
Hence,
(
)
(
Ps max
Pd max
=
B-6
P
maxP s maxP d
)
±
------------------------------------------- -
=
2
<
VSWR
1.05
<
VSWR
1.05
VSWR 1
1.05 1
)
=
---------------------------- -
=
--------------------- -
VSWR
+
1
1.05
Agilent E4416A/E4417A Service Guide
0.0244
=
+
1