Performance Tests
Instrument Accuracy Test
DC
Effective
Calibrator
Power
Setting
Applied
458.31uV
3.1623 uW
1.4493mV
10uW
4.5831mV
31.6228uW
14.493mV
100uW
45.831mV
316.2278uW
144.93mV
1mW
459.12mV
3.1623mW
1.4588V
10mW
4.6872V
31.6228mW
15.588V
100mW
Note
2-14
14. Set the DC calibrator output to the values shown in Table 2-4. For
each setting, set the filter length of the power meter to the
corresponding value, and verify that the power meter reading is
within the limits shown. Refer to step 10 for details on how to
setup the filter length
.
Table 2-4: Instrument Accuracy Results
Power
Meter
Meas
Filter
Unc
Setting
256
0.028 uW
256
0.027 uW
256
0.027 uW
64
0.073 uW
64
0.104 uW
16
0.2 uW
16
0.9 uW
16
1.6 uW
16
4.8 uW
16
18.5 uW
The test limits shown in this table do not take into account the
corresponding measurement uncertainty values.
Lower
Ch A
Limit
Result
3.0977uW
9.904 uW
31.428uW
99.49uW
314.72uW
0.9954mW
3.1477mW
9.954mW
31.477mW
99.54mW
Agilent E4418B/E4419B Service Guide
Ch B
Upper
Result
Limit
3.2269uW
10.096uW
31.819uW
100.51uW
317.74uW
1.0047mW
3.1769mW
10.046mW
31.769mW
100.46mW