Mismatch - Agilent Technologies E4418B Service Manual

Power meters
Hide thumbs Also See for E4418B:
Table of Contents

Advertisement

A normal distribution is assumed, as the 3458A Manual has not specified the type of
distribution used on the measurement.
Both V
the output of the 432A using a 3458A DMM. The approximate values of V
V
scale.
For the 3458A DMM, the specified accuracy for the 10 volt scale is:
8ppm of Reading + 0.5ppm of Range
with;
0.15ppm of Reading + 0.01ppm of Range per °C (temp coefficient Tcal ±5 °C)
± 8.75 ppm of Reading + 0.55ppm of Range
Therefore,
(
U V
comp
For V0 and V1, two measurements are taken, so the uncertainty is equated to the RSS
of using two DMMs;

Mismatch

The Mismatch uncertainty was previously determined as being:
Note
Agilent E4416A/E4417A Service Guide
Measurement Uncertainty Analysis – Power Reference Level Test
and V
values are calculated by subtracting V
0
1
for no power (V
) and for 1mW power (V
ref
0
)
(
)
±
U V
=
=
ref
Voltage measurement
U(Vcomp)
U(V
)
0
U(V
)
1
MsUncertainty magnitude
This is the worst-case uncertainty for mismatch and occurs when the
incident and reflected waves are 180 degrees out of phase.
) are 5V with the DMM on the 10V
1
×
×
8.75
5
0.55
10
---------------------- -
+
----------------------- -
1000000
1000000
49.25 µV
69.65 µV
69.65 µV
(
)
2P s P d
=
Measurement Introduction
from V
as measured at
ref
comp
comp
±
49.25uV
=
Uncertainty
and
B-5