Thermo Scientific GenPure Standard Operating Instructions Manual page 29

Barnstead genpure series water purification system with xcad
Table of Contents

Advertisement

9) Take the cartridge cover off of the GenPure system
10) Take the plugs off from the filter cartridge and save them.
11) Push the filter cartridge in the guide in the system, plug the quick-connect connectors on so
that they audibly click into position and put the cartridge cover back on.
12) Connect the 8 mm o.d. hose that is provided in the connector kit to the feedwater connector
(3) of the GenPure system.
13) Now use the 8 mm o.d. hose to make a pressureless connection between the system
(connector 4) and the drain. The drain to the sewer must be max. Are 1m
above the rinsing water connector of the Unit.
14) Plug the two extension cables with 25 pin plugs in the sockets (1) and (2) of the GenPure
system and screw them tight.
15) If you intend to use the sterile filter (10) that is standardly supplied, screw it in the R 1/4"
female thread outlet of the dispensing valve (9) of the xCAD.
16) Open the feedwater supply.
17) Check all connections for leaks.
Thermo Scientific Barnstead GenPure with xCAD
Ultrapure water system
29