Thermo Scientific GenPure Standard Operating Instructions Manual page 28

Barnstead genpure series water purification system with xcad
Table of Contents

Advertisement

Installing the wall mount
Caution:
The included screws and dowels are only intended to fix the wall mount
to a concrete wall or a solid masonry!
The two 25 pin plugs/extension cables and the 4 mm o.d. and 6 mm o.d. hoses can be led
laterally down the wall or be led out through the wall at the back. There are three possibili-
ties for cut-outs for the cable and hoses (see the next diagram).
Note:
Only one of these cut-outs is required.
2) Hold the wall mount at the wanted position on the wall and mark the four boreholes
for the fixing of the wall mount.
Wand
wall
Wandhalter
wall mount
3) Bore the holes for the 6 mm plugs and make the necessary cut-out.
4) Fix the wall mount in position with the plugs and screws provided.
5) Lead the 25 pin plugs/extension cables and 4 mm and 6 mm o.d. hoses through the wall
to the wall mount and connect the 25 pin plug (1) from the GenPure with the socket (1) on
the xCAD and the 25 pin plug (2) from the GenPure with the socket (2) on the xCAD.
Caution:
Do not interchanging the connections of the 25-pin connector, otherwise no ab-
straction of water possible!
6) Fix the 6 mm und 4 mm o.d. hoses in the hose connectors (5) and (6) of the xCAD.
7) Now carefully position the xCAD on the wall mount and screw it tight to the wall
mount using the four 4x12 mm screws.
8) Place the GenPure system at the planned position.
28
Thermo Scientific Barnstead GenPure with xCAD
min.18
möglicher Ausschnitt in der Wand, für Kabel-
und Schlauchdurchführung nach hinten
possible wall cut- outs for cable
and hose taken out at the back
Ultrapure water system

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents