11.ID Number And GLP Report

The ID number is used to identify the balance when Good Laboratory Practice (GLP) or Good
Manufacturing Practice (GMP) is used.
□ The data output compatible with "GLP/GMP" can be output to a personal computer or printer using the
RS-232C serial interface.
□ The GLP / GMP compliant report includes the balance manufacturer, model, serial number, ID
number, date, time and space for signature. It includes the results and using mass for calibration or
calibration test data.
□ The balance can output the following reports for GLP/ GMP.
・"Calibration report" of the calibration, using the internal mass (Calibration due to changes in
temperature and one-touch calibration.)
・"Calibration report" of the calibration, using an external weight.
・"Calibration test report" of the calibration test, using an external weight.
・"Title block" and "End block" for the weighing data.
□ Calibration and calibration test data can be stored in memory to output several reports at the same
time. Refer to "12. Data Memory" for details.
□ The ID number is used to identify the balance when the balance is used for maintenance manegement.
□ The ID number is maintained in non-volatile memory even if the AC adapter is removed.
□ For details on confirming and setting the time and date. Refer to "10.4. Clock and Calendar
Function".

11-1 Setting The ID Number

Step1 Press and hold the SAMPLE key until ba5fnc of the function table is displayed, then
release the key.
Step2 Press the SAMPLE key several times to display
Step 3 Press the PRINT key. Set the ID number using the following keys.
SAMPLE key ································· To select the digit to change the value.
RE-ZERO key, MODE key ··············· To set the character of the digit selected.
··········································· Refer to the display character set shown below.
PRINT key ···································· To store the new ID number and display
CAL key ······································· To cancel the new ID number and display
Step4 With
Note The display segment of the balance is divided into 4types.
For each segment display, refer to the "Display
correspondence table" in the next page.
displayed, press the CAL key to return to the weighing mode.