A&D GX-203A Instruction Manual page 20

Gx-a series; gx-f series multi-function balance
Hide thumbs Also See for GX-203A:
Table of Contents

Advertisement

Unit setting example
The example below sets the units in the order with g (gram) as the first unit followed by pc (counting
mode).
1
Press and hold the SAMPLE key until ba5fnc of the
function table is displayed, then release the key.
2
Press the SAMPLE key several times to display 1Unit .
3
Press the PRINT key to enter the unit selection mode.
4
Press the RE-ZERO key to specify the unit of g
The stabilization indicator
is specified.
5
Press the SAMPLE key to display 1Unit pC5 .
6
Press the RE-ZERO key to specify the unit of pc
The stabilization indicator
is specified.
7
Press the PRINT key to store the units.
The balance displays
menu item of the function table.
8
Press the CAL key to exit the function table. Then the
balance returns to the weighing mode with g,the unit
selected first.
9
Press the MODE key to switch between g and pc (gpcs).
appears when the unit
appears when the unit
end
and then displays the next
18
Press and hold
Press
several times
Select
Displays the units
sequentially.
Specify
Store