11. ID Number And GLP Report

The ID number is used to identify the balance when Good Laboratory Practice (GLP) is used.
The ID number is maintained in non-volatile memory even if the AC adapter is removed.
The GLP output format is selected at "GLP output (
output to a personal computer or printer using the RS-232C serial interface.
The GLP output format includes the balance manufacturer, model, serial number, ID
number, date, time and space for signature for weighing data, the weight used and results
for calibration or calibration test data.
The balance can output the following reports for GLP.
"Calibration report" of the calibration, using the internal mass (Calibration due to changes
in temperature and one-touch calibration.)
"Calibration report" of the calibration, using an external weight.
"Calibration test report" of the calibration test, using an external weight.
"Title block" and "End block" for the weighing data.
Calibration and calibration test data can be stored in memory to output several reports at the
same time. Refer to "12. Data Memory" for details.
For details on confirming and setting the time and date for the GX-K series. Refer to "10-7.
Clock and Calendar Function".
Caution
GF-K series does not store the calibration report in memory.
GF-K series does not use the time and date function. Use the calendar function
of AD-8121B printer.

11-1. Setting The ID Number

Step 1 Press and hold the
then release the key.
Step 2 Press the
Step 3 Press the
RE-ZERO
SAMPLE
key ......... To store the new ID number and display
PRINT
key ............ To cancel the new ID number and display
CAL
Step 4 With
ba5fnc
Display character set
SAMPLE
key several times to display
SAMPLE
key. Set the ID number using the following keys.
PRINT
key ..... To set the character of the digit selected.
Refer to the display character set shown below.
key ...... To select the digit to change the value.
displayed, press the
)" of the function table and can be
info
key until
of the function table is displayed,
ba5fnc
key to return to the weighing mode.