A&D GX-203A Instruction Manual page 69

Gx-a series; gx-f series multi-function balance
Hide thumbs Also See for GX-203A:
Table of Contents

Advertisement

(2) Using The Statistical Calculation Mode
Entering data for statistical calculation
Use the following keys to operate the statistical calculation mode.
MODE key ................. When the data is entered, moves between the displaying items (weighing mode,
SAMPLE key ............. Turns the minimum weighing value on or off, in the weighing mode.
RE-ZERO key ........... Sets the display to zero in the weighing mode.
PRINT key ................. Outputs the data number and the weight data and includes the weight data to statistical
CAL key ..................... Returns to the weighing mode.
1 Press the RE-ZERO key to set the display to zero.
2 Place the sample on the weighing pan and wait for the
stabilization indicator to turn on.
3 Press the PRINT key to add the data displayed to statistical
calculation. The number of data on the upper left of the
display increases by 1.
4 Repeat steps 1 to 3 for each weighing.
statistical results and data operation) each time the key is pressed.
When no data has been entered, selects the unit.
calculation in the weighing mode. (Output is not in the data format specified in the
function table because of the data number added.)
Outputs the statistical results while the statistical results are displayed. (Output is not in
the data format specified in the function table.)
67
Weighing pan
Sample

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents