Id Number And Glp Report; Main Objective; Setting The Id Number - A&D GX-M Series Instruction Manual

High-capacity precision balances
Hide thumbs Also See for GX-M Series:
Table of Contents

Advertisement

12.ID Number And GLP Report

12-1 Main Objective

The ID number is used to identify the balance when Good Laboratory Practice (GLP) or Good
Manufacturing Practice (GMP) is used.
The data output compatible with "GLP/GMP" can be output to a personal computer or printer using the
RS-232C serial interface.
The GLP / GMP compliant report includes the balance manufacturer, model, serial number, ID
number, date, time and space for signature. It includes the results and using mass for calibration
or calibration test data.
The balance can output the following reports for GLP/ GMP.
・"Calibration report" of the calibration, using the internal mass (Calibration due to changes in
temperature and one-touch calibration.)
・"Calibration report" of the calibration, using an external weight.
・"Calibration test report" of the calibration test, using an external weight.
・"Title block" and "End block" for the weighing data.
□ Calibration and calibration test data can be stored in memory to output several reports at the
same time. Refer to "13. Data Memory" for details.
□ The ID number is used to identify the balance when the balance is used for maintenance management.
□ The ID number is maintained in non-volatile memory even if the AC adapter is removed.
□ For details on confirming and setting the time and date. Refer to "11-7. Clock and Calendar
Function".
□ It is also possible to output the clock data of an external device (such as a printer) without
outputting data of the clock built in to the balance.

12-2 Setting The ID Number

1. Press and hold the SAMPLE key (for 2 seconds) until ba5fnc of the function table is
displayed, then release the key.
2. Press the SAMPLE key several times to display
3. Press the PRINT key. Set the ID number using the following keys.
SAMPLE key ꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏ To select the digit to change the value.
RE-ZERO key,
ꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏ
PRINT key ꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏ To store the new ID number and display
CAL key ꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏ To cancel the new ID number and display
4. With
Note The display segment of the balance is divided into
4types.
For each segment display, refer to the "Display
correspondence table" in the next page.
MODE key ꞏꞏꞏꞏꞏꞏ To set the character of the digit selected.
Refer to the display character set shown below.
displayed, press the CAL key to return to the weighing mode.
id .
72
.
.
7 segment
15 segment
14 segment
11 segment

Advertisement

Table of Contents
loading

Table of Contents