Keysight N3300A User Manual page 91

Dc electronics loads
Table of Contents

Advertisement

B - Performance Test and Calibration
Keysight 3458 or
34401 DVM
Keysight 3458 or
34401 DVM
Iprog Power Source +
& DVM
Note: Use this equipment
only when calibrating
Imon and Iprog gain.
90
A COM
-
I MON
+
Figure B-1. I MON ZERO CALIBRATION
N330xA Module (rear view)
I MON
A11
A COM
A9
EXT PRG
A8
EXT COM
-
A7
If 3 DVM's are not available - move DVM
between I MON, EXT PRG and current shunt as
required by program
Figure B-2. I MON / EXT PRG and CURRENT
CALIBRATION
N330xA Module (top view)
Ribbon
Cable
+
_
SHUNT
Keysight 3458 or
34401 DVM
Location of I MON zero pins
on single-slot modules
IMON_INV
GND_AQS
IMON_INV
GND_AQS
INPUT TERMINALS
MUST BE OPEN
+
_
Power
Source