B - Performance Test and Calibration

Performance Tests

IMON Zero Verification

This test verifies that IMON Zero is within calibration. IMON zero must be in calibration to perform the tests in
this section. If the IMON value is not within calibration, go to the calibration section of this appendix and
calibrate IMON zero, IPROG, current programming and readback and resistance programming and readback.
See figure B-1 for IMON Zero test setup.
Action
1.
Turn off load module and connect DVM to IMON as per figure B-1.
Input terminals must be open ( no power source connected )
2.
Turn on Load – press [ Recall } – scroll to *RST – press [Enter ]
3.
Read IMON voltage from DVM.
4.
If IMON in step 3 in not within specification then calibrate IMON
zero, IPROG, Current mode and Resistance mode.

CC Mode Tests

These tests verify that the module operates in the CC mode and that IMON, current programming, and readback
to the front panel display are within specification. Values read back over the GPIB are the same as those
displayed on the front panel. To read back the current value via the GPIB, use MEAS:CURR:ACDC?
Current Programming and Measurement Accuracy Test
This test verifies that the current programming and measurement accuracy are within specification. If the test
readings are out of tolerance the module may require calibration. If the test readings significantly disagree with
the specified values or no readings can be obtained go to the Turn-On Checkout procedure to verify module
operation.
Make a copy of the module test card for the model to be tested to record test values.
1.
Connect the Electronic Load, power source, DVM and current shunt as shown in figure B-2.
2.
Turn on the Electronic Load and press [ Recall ]. Scroll until display reads *RST and press [ Enter ].
3.
Turn on the power source. Set the power source voltage and current to the values listed in the following table.
Power Source Voltage Setting
Power Source Current Setting
4.
Checking high current range, low current point. Press [ CURR ] [ 1 ] then [ Enter ]. Wait 10 seconds then record the
actual input current ( DVM reading / current shunt resistance ) and front panel current reading on the test record card
under high current range low current.
5.
Checking high current range, high current point. Press [ CURR ] [ select current from following table] [Enter ].
N3302A
N3303A
30A
10A
6.
Wait 30 seconds, then record the actual input current ( DVM reading / current shunt resistance ) and front panel current
reading on the test record card under high current range high current.
78
N3302A
N3303A
5V
20V
33A
12A
N3304A
N3305A
60A
60A
Normal Result
Voltage reading less than 0.166mV for
models N3302A - N3305A, and N3307A.
Voltage reading less than 0.320mV
for model N3306A
N3304A
N3305A
5V
8V
61A
61A
N3306A
N3307A
120A
30A
N3306A
N3307A
5V
5V
122A
33A