B - Performance Test and Calibration
Power Source Voltage Setting
Power Source Current Setting
5.
Checking high voltage range, low voltage point. Press [ VOLT ] [ 3 ] [ Enter ]. Wait 10 seconds, then record actual
input voltage (DVM reading) and front panel readings on test record card under high voltage range, low voltage.
6.
Checking high voltage range, high voltage point. Press [ VOLT] [select voltage from the following table] [Enter].
N3302A
N3303A
60V
240V
7.
Wait 10 seconds, then record the actual input voltage (DVM reading) and front panel voltage reading on the test record
card under high voltage range, high voltage.
8.
Checking low voltage range, low voltage point. Press [ VOLT ] [ 3 ] [ Enter ]. Press [ VOLT ]. Scroll until display
reads V:RANG, press [ 3 ] [ Enter ]. Press shift key then [ Sense ]. Scroll until display reads S:V:RNG, press [ 3 ]
[Enter]. Wait 10 seconds then record actual input voltage (DVM reading) and front panel readings on test record card
under low voltage range, low voltage.
9.
Checking low voltage range, high voltage point. Press [ VOLT ] [ select voltage from following table ] [ Enter ].
N3302A
N3303A
6V
24V
10. Wait 10 seconds then record actual input voltage (DVM reading) and front panel readings on test record card under
low voltage range, low voltage.
CV Mode Regulation Test
This test verifies the Input Voltage remains within specification when the input current is changed from a low
current to rated current.
1.
Connect Electronic Load, power source and DVM as shown in figure B-3. Be sure + sense and – sense are connected
to the +/- input terminals and the remote/local button is in the remote position.
2.
Turn on the Electronic Load and press [ Recall ]. Scroll until display reads *RST and press [ Enter ].
3.
Press [ Func ]. Scroll until display reads FUNC VOLT, then press [ Enter ].
4.
Press [ Volt ] [ select voltage from following table ] [ Enter ].
N3302A
N3303A
5V
25V
5.
Turn on power source. Set power source voltage and current to values listed in following table.
Power Source Voltage Setting
Power Source Current Setting
6.
Wait 10 seconds and record input voltage reading.
7.
Reset power source current level to value listed in following table.
Power Source Current Setting
8.
Wait 10 seconds and record input voltage reading.
9.
Subtract reading in step '8' from reading in step '6'. Absolute value of difference should be less than specification.
80
N3302A
N3303A
61V
246V
2A
0.6A
N3304A
N3305A
60V
150V
N3304A
N3305A
6V
15V
N3304A
N3305A
5V
8.3V
N3302A
N3303A
6V
30V
1A
0.4A
N3302A
N3303A
30A
10A
N3304A
N3305A
61V
152V
3A
2A
N3306A
N3307A
60V
150V
N3306A
N3307A
6V
15V
N3306A
N3307A
5V
8.3V
N3304A
N3305A
6.5V
10V
1A
1A
N3304A
N3305A
60A
60A
N3306A
N3307A
61V
152V
3A
2A
N3306A
N3307A
6.6V
10V
1A
1A
N3306A
N3307A
120A
30A