B - Performance Test and Calibration
1620
CLEAR SCREEN
1630
!
1640
PRINT TABXY(10,12);"Set power source to voltage and current settings from"
1650
PRINT TABXY(10,13);"Table B-1 for resistance range 4 - Point P1"
1660
DISP "Press CONT when ready to calibrate resistance range 4"
1670
PAUSE
1680
CLEAR SCREEN
1690
PRINT TABXY(9,15);"CALIBRATING RESISTANCE RANGE 4 - P1 - 3 SECOND WAIT"
1700
OUTPUT @Ld;"INP:STAT ON"
1710
OUTPUT @Ld;"RES:RANG 2000"!< resistance range 4 variable from table B-1
1720
OUTPUT @Ld;"CAL:LEV P1"
1730
WAIT 3
1740
INPUT "Measure voltage across current shunt - enter value in volts",Vp1
1750
Ip1=Vp1/Rshunt4
1760
INPUT "Measure voltage across +/- sense terminals and enter value in
volts",Vt1
1770
Rp1=Vt1/Ip1
1780
OUTPUT @Ld;"CAL:DATA";Rp1
1790
!
1800
CLEAR SCREEN
1810
PRINT TABXY(10,12);"Set power source to voltage and current settings from
Table B-2"
1820
PRINT TABXY(10,13);"for resistance range 4 - Point P2"
1830
DISP "Press Cont to continue resistance range 4 calibration"
1840
PAUSE
1850
OUTPUT @Ld;"CAL:LEV P2"
1860
CLEAR SCREEN
1870
PRINT TABXY(9,15);"CALIBRATING RESISTANCE RANGE 4 - P2 - 3 SECOND
1880
WAIT 3
1890
INPUT "Measure voltage across current shunt - enter value in volts",Vp2
1900
Ip2=Vp2/Rshunt4
1910
INPUT "Measure voltage across +/- sense terminals and enter value in
volts",Vt2
1920
Rp2=Vt2/Ip2
1930
OUTPUT @Ld;"CAL:DATA";Rp2
1940
OUTPUT @Ld;"INP:STAT OFF"
1950
!
1960
OUTPUT @Ld;"CAL:SAVE"
1970
OUTPUT @Ld;"CAL:STAT OFF"
1980
!
1990
CLEAR SCREEN
2000
PRINT TABXY(10,10);"Resistance calibration adjustments complete"
2010
PRINT TABXY(10,12);"Verify Resistance Adjustments"
2020
END