Module Specific Error Messages

These errors are detector specific.

Internal temperature exceeded lower limit

Error ID: 10
Air temperature inside the instrument is ≤10 °C
Probable cause
1

Internal temperature exceeded upper limit

Error ID: 11
Air temperature inside the instrument is >40 °C
Probable cause
1
Agilent 1290 Infinity II ELSD User Manual
The environmental temperature is outside
the specified operating limits of the
instrument
The environmental temperature is outside
the specified operating limits of the
instrument
Error Information

Module Specific Error Messages

Suggested actions
• Increase ambient temperature where
detector is located
• Move the detector to a warmer location
Suggested actions
• Decrease ambient temperature where
detector is located
• Move the detector to a cooler location