LAN Configuration
TCP/IP Parameter Configuration

Configuration Using Instrument Front Panel

Configuration of the TCP/IP parameters via the front panel is the
recommended method for setting up the module's LAN interface. On- board
configuration is only possible during the module's start- up procedure.
1 Power on the detector, and press any front key when the following
If no key is pressed within 5 seconds, the ELSD will continue its start-up routine. The
N O T E
previously stored TCP/IP parameters will be used.
2 Select LAN using left/right arrow keys and press AZ/Stop to confirm.
3 Select either Static or DHCP (automatic) IP address assignment and
4 If Dynamic IP address is selected, the instrument will try to obtain an
5 If Static IP address is selected, the instrument will display the stored
The detector will use the default TCP/IP settings if no IP address has been programmed
N O T E
previously (see
6 To enter a new IP address, use the up/down arrow keys.
7 To enter a new Subnet Mask use the up/down arrow keys.
8 To enter a new Gateway address use the up/down arrow keys.
9 When all values are populated press AZ/Stop to accept.
The instrument will return an INVALID network setting message if the IP address and
N O T E
Gateway combination are incompatible.
36
message is displayed
press AZ/Stop to confirm.
IP address from the network.
When successful the detector will continue to the main menu screen.
static TCP/IP settings.
Table 5
on page 35).
"Press a key to enter communication
Agilent 1290 Infinity II ELSD User Manual
setup".