Spiky Peak Tops But Flat Baseline; Large Baseline Offset - Agilent Technologies 1290 Infinity II User Manual

Evaporative light scattering detector
Hide thumbs Also See for 1290 Infinity II:
Table of Contents

Advertisement

7
Troubleshooting and Diagnostics
General Problems

Spiky peak tops but flat baseline

Probable cause
1
2
3
4
5
6

Large Baseline offset

Probable cause
1
2
3
4
78
Inconsistent nebulization
Incorrect gas being used
Poor regulation of inlet gases
Insufficient smoothing
Sample precipitation during nebulization
Inconsistent pump flow rates
Inefficient evaporation
High concentration of non-volatile buffer or
stabiliser
Contaminated diffuser
Optics Heater failed
Suggested actions
Nitrogen is the recommended gas - others can be
used but may not nebulize as efficiently
Change gas to nitrogen or evaluate different
nitrogen sources
• If using bottled gas, check that gas regulator
is functioning correctly and giving consistent
flow
• Alternatively, use a pulse dampener
The broader the peaks, the higher the smoothing
value is required. Increase smoothing to 50 for
Flash separations
• Reduce sample concentration or inject same
loading in larger volume
• Check sample solubility in mobile phase
eluents
See remedy for pump pulsation in the baseline
noise section
Suggested actions
Increase the evaporator temperature and/or gas
flow
Use a lower concentration of stabiliser,
unstabilised solvent or a more volatile buffer
(ammonium acetate or ammonium formate)
Perform cleaning procedure
Refer to local distributor or Agilent Technologies
Agilent 1290 Infinity II ELSD User Manual

Advertisement

Table of Contents
loading

This manual is also suitable for:

G7102a

Table of Contents