Unit Selection; Frequency Team Selection; Alarm Limit Setting And Alarm Start - Lutron Electronics EMF-819 Operation Manuals

50 mhz to 3 ghz radio frequency radiation meters, electromagnetic field strength measurement, 3 axis rf electromagnetic field meter
Table of Contents

Advertisement

7-3 Unit Selection

After inserting the " Probe Memory Card " ( 5-10, Fig. 1 ) and
connecting " Probe Plug " ( 5-18, Fig. 1 ) into the " Probe
Input Socket " ( 5-9, Fig. 1 ), use " Unit Button " ( 5-6, Fig. 1)
to select the " V/m, W/m^2, mW/cm^2 ".
7-4 Frequency Teams Selection
1) Use the " Freq. Team Button " ( 5-5, Fig. 1 ) to select
the proper frequency team.
2) The default selection is " Normal ", however if the
measurement frequency is microwave or its frequency is
near " 2.45 GHz ", it should select to " 2.45 GHz "
to get the high precision.
3) If user don't know the exact information, just select to
the " Normal ".

7-5 Alarm limit setting and Alarm start

1) Press the "Alarm Button " ( 5-8, Fig. 1) once to start
the alarm function, while the " units " symbol will
flash on the display.
2) Press the " Alarm Button " ( 5-8, Fig. 1 ) once again
to stop the alarm function.
3) Press the " Alarm Button " ( 5-8, Fig. 1 )
continuously and over 2 seconds, the "Alarm low
limit value " will indicate on the lower display, use
the "
,
▼ ▲
low limit value.
4) Press the " Enter Button " ( 5-4, Fig. 1 ) will save the
low limit value and going to adjust the " High limit
value, procedures will similar as above.
Buttons " to adjust the desiring
10