Re-Calibration – 1 Point EDIT and START
To perform a 1-point re-calibration, apply a pressure between 50%
and 100% of Full Scale and then follow these steps:
Keystroke
1. With unit OFF, press and
hold the MIN/MAX key,
turn the unit on by
pressing the ON/OFF
key, then release
MIN/MAX.
2. Press the  arrow key
until "START" is
displayed on the bottom
line.
3. Press the PRGM key.
4. Press the PRGM key.
5. Press the  or  arrow
keys to edit the selected
digit. Use the  or 
arrow keys to change the
cursor position. Value
entered must be 50-
100% of FS.
6. Press the  arrow key
while on the right most
digit to proceed.
7. Apply the input pressure
indicated using an
appropriate reference
standard; press PRGM
key.
8. Press the  arrow key.
28 of 36
Display
Top line reads "RE-CAL".
Bottom line reads "EDIT".
Top line reads "RE-CAL".
Bottom line reads "START".
Top line reads "RE-CAL
START".
Bottom line reads "1-POINT".
Top line reads "CAL POINT"
and bottom line displays the
cal point value.
Bottom line displays the cal
point value. The cursor flashes
at the first position while the
value is changed, then moves
to the right position when the
right arrow key is pressed.
Top line reads "APPLY:"
Bottom line displays the "CAL
POINT" value.
Top line reads "RE-CAL".
Bottom line reads "START",
Manometer has been
recalibrated.
Returns to Measure Mode