Data Logging - Omega HHP360 Series User Manual

Precision absolute manometer
Table of Contents

Advertisement

Data Logging

Data Logging can be used to record pressure measurements. Two
record modes are supported: automatic and manual. In automatic
mode, a pressure value is captured every 5 seconds for 20 minutes,
resulting in 240 stored values. In manual mode, a pressure value is
captured each time the PRGM key is pressed up to 240 values. The
data collected during a logging session can be viewed upon
completion.
Keystroke
1. From the Measure Mode
press the PRGM key.
2. Press the  arrow key
five (5) times.
3. Press the PRGM key.
4. Press the PRGM key.
5. Press the PRGM key at
AUTO to start automatic
logging or at MANUAL
to start manual logging
mode.
6. To stop recording values
at any time, press the
key.
7. To access recorded
values, press the
key.
8. To view recorded values,
press the PRGM key.
9. Press the
key 3 times.
The "Auto Shut-Off" timer is disabled for Data Logging sessions.
Be sure to end the session to re-enable the Auto Shut-Off timer.
25 of 36
Display
Top line reads "PROGRAM
MODE" and bottom line
reads "UNITS SELECT".
Bottom line reads "DATA
LOGGING".
Top line reads "DATA
LOGGING" and bottom line
reads "RECORD".
Top line reads "RECORD
MODE" and bottom line
reads "AUTO" or
"MANUAL".
Top line reads
"RECORDING
X" and
bottom line reads "XX.XX
UNITS". AUTO records
value every 5 seconds.
Manual records value each
time PRGM key is pressed.
Top line reads "DATA
LOGGING" and bottom line
reads "RECORD".
Top line reads "DATA
LOGGING" and bottom line
reads "VIEW".
Top line reads "DATA LOG:
1" and bottom line displays
the value. Continue pressing
the ▲ key to view all values.
Returns to Measure Mode.