Specifications - Omega HHP360 Series User Manual

Precision absolute manometer
Table of Contents

Advertisement

Specifications

Type and Range and Display Resolution:
HHP360-A: 0 - 17.403 psia (0 - 900.00 mmHg)
HHP360-B: 0 - 38.674 psia (0 - 2000.0 mmHg)
Accuracy:
HHP360-A: 0017: ±0.02 % F.S. (F.S. = 900 mm Hg)
HHP360-B: 0038: ±0.015 % F.S.* from 0-1000 mm Hg
±0.025% F.S.* from 1000-2000 mm Hg
*F.S. = 2000 mm Hg Absolute
Includes the combined effects of temperature, linearity,
repeatability, hysteresis and resolution.
Warm up time = 5 minutes.
Temperature:
Storage = -40°C to +60°C (-40°F to +140°F)
Operating = -5°C to +50°C (23°F to +122°F)
Media Compatibility:
AI: Absolute pressure sensors for use with gases and liquids
compatible with 316L SS
Pressure Limits:
AI units: 77 PSIA (4000 mm Hg Abs)
Connection:
1/8" female NPT, 316L SS. P1 is the pressure connection.
P2 is not accessible (factory plugged with metal disc)
User must use a wrench on the pressure manifold when
installing user's 1/8" NPT fitting. Do not tighten the
fitting without using a wrench on the pressure
manifold. Failure to use a wrench on the manifold will
damage the plastic enclosure and void warranty No
torque should be applied to the manifold with respect
to plastic enclosure.
Battery Operation:
>100 hours continuous use, 1 year shelf life, auto power off
programmable at Disabled, 10, 20, 30, 60 or 90 minutes
Enclosure: (6.9" x 3.8" x 2.3") Polycarbonate, Permanently
Static Dissipative, ESD Protection
Enclosure with Boot: (7.2" x 4.2" x 2.5")