Maintenance

Calibration

To maintain high accuracy, the instrument should be calibrated once every year.
Calibration services are also available from your nearest representative. The following
example of calibration using the standard equipment recommended in the section
"Selecting the Standards" is described.
Selecting the Standards
Generation Function
Item to be
calibrated
DCV
DCA
Measurement Function
Item to be
calibrated
DCV
DCA
Required Environment during Calibration
Ambient temperature:
Relative humidity:
Power supply voltage:
Power supply frequency: (50/60 Hz)±1 Hz
Warm-up time:
Point to Note during Calibration
Make sure that the AC outlet to which this instrument is to be connected is a 3-slot
outlet with a grounding terminal.
Standard
Calibration
Measurement
Range
Range
Digital
100 mV
Max. 110 mV
multimeter
1 V
Max. 1.1 V
10 V
Max. 11 V
Digital
20 mA
22 mA
multimeter
500 Ω
Digital
Max. 3 V
multimeter
5 kΩ
50 kΩ
Standard
Calibration
Generation
Range
Range
Standard DC 500 mV
500 mV
voltage
5 V
5 V
generator
35 V
35 V
Standard DC 20 mA
20 mA
current
100 mA
100 mA
generator
500 Ω
500 Ω
Precision-
resistance
5 kΩ
5 kΩ
digital
50 kΩ
50 kΩ
multimeter
23±1˚C
45% to 75% RH
100 V±5%
Before starting calibration, the standard equipment must be
warmed up for the specified time, and the instrument must be
warmed up for at least 4 hours.
Accuracy
±(10 ppm+1 µV)
±(10 ppm+5 µV)
±(10 ppm+50 µV)
±(50 ppm+0.4 µA)
±(10 ppm+1 mV)
Accuracy
±(20 ppm+5 µV)
±(20 ppm+50 µV)
±(20 ppm+300 µV)
±(60 ppm+0.4 µA)
±(40 ppm+1 µA)
±50 ppm(5 ppm/˚C)
41