Precision absolute manometer
Table of Contents

Advertisement

2.
To zero the manometer using Factory Zero, start with the
unit turn ON and in Measure Mode and use the following
keystroke sequence:
Keystroke
1.
Press MIN/MAX
and TARE keys at
the same time. (See
figure 1 above.)
2.
Press ▲ or ▼ arrow
key until desired
zero function is
shown on the bottom
line.
3.
Press the PRGM
key.
NOTE: The HHP360 can be zeroed only if the new applied zero is
within ± 1% FS of the original factory calibration zero. If outside
this limit a "ZERO RANGE ERROR" message appears and the
manometer will not zero. Contact the factory for support in this
case.
9 of 36
Display
Top line of display reads
"ZEROING SOURCE:" Bottom
line of display reads "REF TO
ABS ZERO"
Bottom line of display
reads "FACTORY ZERO"
Zeroing is complete when unit
returns to Measure Mode.