Leak Test

The Leak Test feature allows the user to determine the leak rate in
the pneumatic system being monitored. Once configured, Leak
Test monitors the measured pressure over time and displays the
leak rate in the pressure units per minute at the conclusion of the
test. The maximum configurable leak test period is 1440 min (1
day). Pressing any key during the leak test will abort the test.
To enable Leak Test follow these steps:
Keystroke
1. From the Measure
Mode press the PRGM
key.
2. Press the  arrow key
twice.
3. Press the PRGM key.
4. Press the PRGM key.
5. Use the  ,  & 
keys to input test period
6. Press the PRGM key.
7. Press the  arrow key
once.
8. Press the PRGM key.
9. Press the PRGM key to
exit the Leak Test.
The "Auto Shut-Off" timer is disabled for Leak Test sessions. Be
sure to end the session to re-enable the Auto Shut-Off timer.
26 of 36
Display
Top line reads "PROGRAM
MODE" and bottom line reads
"UNITS SELECT".
Bottom line reads "LEAK
TEST"
Top line reads "LEAK TEST"
and bottom line reads
"CONFIGURE".
Top line reads "Leak Test
Period" & bottom "X.X MIN".
Bottom line reads desired
period; Ex. " 20.0 MIN".
Top line reads "LEAK TEST"
and bottom line reads
"CONFIGURE".
Top line reads "LEAK TEST"
and bottom line reads "PRGM
TO START".
Top line displays MIN/MAX
pressure values at left/right.
Bottom line reads the current
pressure value and units.
At end of test period, top line
displays the leak rate in units
per minute. Bottom line shows
the current pressure reading.
Returns to Measure Mode.