BEFORE COMMENCING OPERATION

Sterilizing of chamber and attachments

Before first start-up of the unit, the chamber and internal attachments should be cleaned and sterilized as
follows.
Note:
Take care not to damage the UV lamp or water level sensor at the time of removal or replacement of
attachments.
Do not clean the inside of the unit with a solution of disodium chlorate or other halogen-based solution
because this may cause corrosion of metal surfaces.
1. Take out all attachment such as trays, humidifying pans, etc, from the chamber. Refer to page 27 and
28.
2. Clean all of the attachments with neutral detergent and then wash out the detergent with distilled water.
3. Wipe the attachments with a gauze containing alcohol for sterilization and then wipe off with a dry
gauze.
4. Wipe the inside wall of the chamber with a gauze containing alcohol for sterilization and then wipe off
with a dry gauze.
5. Replace all attachments in the chamber.
Note:
Always insert the fan on the motor shaft surely. Improper insertion may cause poor performance.
6. Fill the humidifying pan with sterile distilled water.