Performance specification
Model
Ambient
temperature
humidity range
Temperature control range in
the unit
Temperature distribution in
the unit
Temperature variation in the
unit
Temperature recovery period
CO
level range
2
CO
level distribution range
2
CO
level recovery period
2
Internal humidity
Internal
humidity
recovery
period
Supplied CO
gas pressure
2
Power supply
Total power consumption
Total current
Quantity of radiation
Test condition:
Set point: 37 C
CO
set point: 5.0%
2
No load
Ambient temperature: 20 C
Ambient humidity: Approx.45%
Based on the measuring method on validation service manual.
MCO-18AIC
Temperature: 5 C 35 C
and
Humidity: Less than 80%RH
(In ambient temperature is lower than 15 C, original
performance cannot be always obtained)
Ambient temperature+5 C +50 C
(Settable range:0 C 50 C)
0.25 C
0.1 C
25 minutes or less
(until the internal temperature is recovered to 36.5 C after
door opened for 30 seconds)
0 20%
0.15%
10 minutes or less
(until the CO
level is recovered to 4.5% after door opened
2
for 30 seconds.)
95 5%RH
30 minutes or less
(until the internal humidity is recovered to 90%RH after door
opened for 30 seconds.)
0.03MpaG (0.3kg/cm
Single phase, local voltage
310W
110-120VAC: Maximum 2.8A
220-240VAC: Maximum 1.4A
Maximum 1120kJ/h
3
2
G) when the gas supplied